Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
MEASURED PARAMETERS
PRESSURE
Range: 0-250mmHg
Accuracy: ±2mmHg
Precision: ±1mmHg
Resolution: 1mmHg
Zero: Automatic
PULSE VOLUME (AC PRESSURE)
Range: ±10mmHg
Accuracy: ±0.1mmHg
Precision: ±0.1mmHg
Resolution: 0.01mmHg
Bandwidth: 1Hz-10Hz (-3dB)
Sampling rate (all parameters): 40Hz
ANALOGUE OUTPUTS
PRESSURE
Range: 0-3V
Transfer function: 1V = 100mmHg
Resolution: 12-Bit
PULSE VOLUME
Range: 0-5V
DC Offset: 2.5V
Transfer function: ±2.5V = ±10mmHg
Resolution: 12-Bit
Connectors: 50 Ohm BNC sockets
OUTPUTS
Digital LCD screen display
USB interface for connection to PC
PUMP
Inflation range: 20-250mmHg
Inflation rate: Inflates a leg cuff to 200mmHg in less than 10 seconds, arm cuff in less than 5 seconds.
Deflation rate: Linear, 1mmHg/s – 10mmHg/s from 250-40mmHg for all cuff sizes.
GENERAL
Universal voltage switch-mode power supply range 100-230V at 30VA, 50 to 60 Hz
Dimensions W H D mm: 235 x 80 x 200
Weight kgs: 2.0
STORAGE AND TRANSPORT CONDITIONS
Temperature 0-45°C
Humidity 0-80% RH
Atmospheric pressure 500-1060hPa
Medical devices classification: Class IIa Active device for diagnosis
Type of protection against electric shock - Class 1
Degree of protection against electric shock - Type BF
Degree of protection against ingress of liquid - IPXO (not protected)
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with nitrous oxide.
Mode of operation: Continuous.