Moor Instruments are committed to product development. We reserve the right to change the specifications below without notice.
OPTICS
Laser Safety Classification
Class 3R per IEC 60825-1:2014
Output power 20mW max.
Wavelength 785nm ± 10nm
Angular spread of laser light from probe tip 26°
All specifications include cumulative measured uncertainties and expected increases in values after manufacture.
PERFUSION MEASUREMENTS
FLUX
Accuracy: ± 10%, relative to Moor Instruments ‘standard’ moorVMS-LDF1-HP
Precision: ± 3% of measurement value
Range: 0-1000 PU
CONC
Accuracy: ± 10%
Precision: ± 5% of measurement value
Range: 0-1000AU
DC
Accuracy: ± 10%
Precision: ± 3%
Range: 0-1000AU
TEMPERATURE MEASUREMENT
Range 5-50°C
Resolution 0.1°C, accuracy ± 0.3°C
SIGNAL PROCESSING
Bandwidth: high pass 20Hz
Low pass selectable 3KHz*, 15KHz, 22KHz*
Output time constants: 0.1s, 0.5s, 1.0s, 3.0s and unfiltered
Automatic gain control and zeroing
*Feature only available whilst using the moorVMS-PC Windows™ software
OUTPUTS
Digital LCD screen display
40Hz maximum data rate
0-1000 arbitrary perfusion units
USB interface for connection to PC
2 analogue outputs per channel, BNC sockets 0-5v
INTERNAL SOFTWARE FEATURES
Flash memory programmable from PC
Output scale adjustment
Filter bandwidth selection
GENERAL
Universal voltage switch-mode power supply range 100-230v at 30VA, 50 to 60 Hz
Dimensions W H D mm
moorVMS-LDF1-HP 235 x 60 x 200
Weight kgs
moorVMS-LDF1-HP 1.4
Mode of operation: continuous.
STORAGE AND TRANSPORT CONDITIONS
Temperature 0-45°C
Humidity 0-80% RH
Atmospheric pressure 500-1060hPa
Type of protection against electric shock - Class 1
Degree of protection against electric shock - Type BF
Degree of protection against ingress of liquid - IPXO (not protected)
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with nitrous oxide.